How to Check Which Gases a PID Sensor Can Detect
Step 1: Look Up the Reference Table
The simplest way is to look up the ionization energy and CF value of the VOC gas in the reference table. Search for the gas by its molecular formula or name to locate it in the table. Check the commonly used 10.6 eV column: if the CF value is 0, it means the gas cannot be ionized by 10.6 eV energy. If it is not 0, that value is the CF value of the gas — a coefficient that helps us calculate the corresponding concentration.
Although the table lists several hundred common substances, the variety of VOCs goes far beyond that. If the substance you need cannot be found in this table, proceed to Step 2.
Step 2: Search the Website
https://webbook.nist.gov/chemistry/form-ser/
This website covers most of the substances that have been studied and tested. To use it:
1. Copy the link above and open it in any web browser, then enter the molecular formula of the substance you want to check — for benzene, C6H6 — in the search box on the page, and click Search at the bottom left;
2. In the results, find a corresponding molecular structure of benzene, Benzene C6H6, and click to open it;
3. On the page, find Gas phase ion energetics data and click to open it, then look up the ionization energy value of that molecular structure;
4. As shown, the ionization energy of benzene is 9.24 eV. The UV lamp we commonly use is 10.6 eV. Since the energy of the UV lamp is greater than the energy required to ionize the substance, the gas can be ionized — therefore the PID sensor can respond to it
.
Conversely: for example, the IE (eV) of carbon monoxide (CO) is 14.01 eV, so it can never be ionized by a 10.6 eV PID sensor.
With these two steps, you can identify whether the vast majority of common substances can be detected by a PID sensor.